Eversense CGM User Guide
Alert Descriptions and Actions
(continued)
Alerts
App Displa
y
Description
Calibrate Now
Appears to alert you that your calibration is due.
If you do not calibrate within 4 hours, glucose values
will no longer be displayed.
Sensor Check
Appears once when the system’s internal checks
detect instability with the sensor which requires a
return to calibration Initialization Phase.
A
ctions
Tap
Calibrate
to enter a calibration value.
In the Initialization Phase, you must perform
4 fingerstick calibration tests spaced 2 - 12 hours
apart. Display of glucose readings will resume after
the 2nd successful fingerstick calibration test.