Eversense CGM User Guide
Electrical and Safety Standards
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The transmitter is intended for use in the electromagnetic environment specified in the next table. The customer or
the user of the transmitter should ensure that it is used in such an environment.
Immunity
Test
Immunity
Test
Transmitter
Compliance
Level
Electromagnetic
Environment
Guidance
Electrostatic Discharge
(ESD) IEC 61000-4-2
± 8 kV Contact
± 15 kV Air
± 8 kV Contact
± 15 kV Air
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Power Frequency
(110VAC/60Hz,
230VAC/50 Hz)
Magnetic Field IEC
61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
Transmitter Electromagnetic Immunity Specifications