19
179
Eversense CGM User Guide
Table 10 – Eversense Agreement when “LO” or “HI” is Displayed in the PRECISION Study
Glucose values below 40 mg/dL or above 400 mg/dL are shown on the mobile app as “LO” or “HI”. The table below
shows, as a percentage, agreement of sensor glucose values when outside the display range of 40 mg/dL to 400 mg/dL.
Reference mg/dL
CGM
Readings
CGM-Ref
pairs
< 50
< 60
< 70
< 80
> 80
Total
‘LO’
n
6
7
9
9
0
9
%
66.7%
77.8%
100%
100%
0%
100%
Reference mg/dL
CGM
Readings
CGM-Ref
pairs
> 340
> 320
> 280
> 240
< 240
Total
‘HI’
n
359
383
399
404
1
405
%
88.6%
94.6%
98.5%
99.8%
0.2%
100%