Eversense CGM User Guide
Alert Descriptions and Actions
(continued)
Alerts
App Displa
y
Description
Calibration Past Due
Appears when your system is past due for
calibration.
No glucose readings can be displayed until
calibration is performed.
Calibration Expired
Appears when a calibration has not been performed
in 24 hours. The system returns to the Initialization
Phase.
No glucose reading is displayed until calibration is
performed.
A
ctions
Perform a fingerstick calibration in order to resume
displaying glucose values.
In the Initialization Phase, you must perform
4 fingerstick calibration tests spaced 2 - 12 hours
apart. Display of glucose readings will resume after
the 2nd successful fingerstick calibration test.