 Power/Set Button 

2

 Display

3

 Right Arrow and Left Arrow Buttons 

  

4

 Test Strip Slot

5

 Battery Door

6

 Infrared (IR) Window

Meter Overview

Start Here

Quick Start Guide

If any of the error messages below appear on the display at any time, 

refer immediately to the User’s Manual, Chapter 5, Display and Error 

Messages. 

E-1

 The test strip may be damaged or not properly inserted. Remove and 

reinsert the test strip, or replace it if damaged. 

E-3

 Your blood glucose may be extremely high or a meter or a test strip 

error has occurred. If your test result matches how you feel, contact your 

healthcare professional immediately. If your test result does not match 

how you feel, repeat the blood glucose test. See the User’s Manual, 

Chapter 2, Unusual Blood Glucose Results. If the second test result does 

not match how you feel, perform a control test with the control solution 

and a new test strip. If the control result is within the acceptable range, 

review the proper testing procedure and repeat the blood glucose test 

with a new test strip. If the E-3 code still appears for your blood glucose 

test, your blood glucose result may be extremely high and above the 

system’s reading range. 

Contact your healthcare professional 

immediately.

 If the control result is not within the acceptable range, see 

the User’s Manual, Chapter 4, Understanding Out-of-Range Control 

Results. 

E-4

 Not enough blood or control solution was drawn into the test strip for 

measurement or was applied after the test had started. Discard the test 

strip and repeat the blood glucose or control test. 

E-6

 Blood or control solution was applied to the test strip before the 

flashing drop symbol appeared on the display. Discard the test strip and 

repeat the blood glucose or control test. 

E-7

 An electronic error occurred, or in rare cases, a used test strip was 

removed and reinserted. Turn the meter off and on, or take the battery 

out for 20 seconds and reinsert it. Perform a blood glucose or control 

test. 

E-8

 The temperature is above or below the proper range for the system. 

Refer to the test strip package insert for system operating conditions. 

Move to an area with the appropriate conditions, wait 5 minutes, and 

repeat the blood glucose or control test. Do not artificially heat or cool 

the meter. 

E-9

 The battery is almost out of power. Change the battery now. If the 

message reappears after the battery has been replaced, remove the 

battery again, press any meter button, then reinsert the battery.

E-10

 The time and date settings may be incorrect. Make sure the time 

and date are correct and adjust, if necessary. 

Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim, Germany
www.accu-chek.com

Wash and dry your hands before 

you perform a blood glucose test.

1

Check the use by date on the test 

strip container. Do not use test 

strips past the use by date.

2

Remove a test strip from the test 

strip container. Close the cap 

tightly.

3

Insert the test strip into the meter 

in the direction of the arrows until 

the meter beeps.

4

A flashing blood drop symbol 

appears.

Performing a Blood Glucose Test

Hold the lancing device firmly 

against the edge of your fingertip 

and press the release button to 

prick your finger.

Touch the end of the test strip to 

the blood drop. Do not put blood on 

top of the test strip.

The meter beeps and   flashes 

when there is enough blood in the 

test strip.

The test result appears on the 

display after 5 seconds.

Error Messages

5

6

NOTE

Contact Roche if the error message continues. 

ACCU-CHEK, ACCU-CHEK PERFORMA, and FASTCLIX are 

trademarks of Roche.
© 2013 Roche Diagnostics
07054238001(01)-0713

LAST UPDATE: 2013-04

or

1

2

3

4

5

6

1

 WarNiNG

•  This Quick Start Guide does not replace the 

User’s Manual for your Accu-Chek Performa 

blood glucose meter. The User’s Manual 

contains important handling instructions and 

additional information.

•  The lancing device is intended for personal 

use only.

Familiarize yourself with the 

Accu-Chek FastClix lancing 

device.

1

Get the lancet drum from the 

pocket in the carry case.

2

NOTE

Each lancet drum contains 6 sterile 

lancets.

Remove the cap by pulling it 

straight off. Do not twist the cap.

3

Insert the lancet drum, white end 

first, into the lancing device. Push 

the lancet drum all the way in until 

it clicks firmly into place.

4

NOTE

Do not remove the lancet drum 

until all 6 lancets are used since 

the drum cannot be reused once it 

has been removed from the 

lancing device.

Replace the cap on the lancing 

device. Make sure the notch on the 

cap lines up with the notch on the 

lancing device.

5

Adjust the lancet depth to a 

comfortable penetration level. Start 

at 2. For tougher skin, dial to a 

higher number.

Using the lancing device: By 

pressing the release button, the 

lancing device primes and pricks 

with 1 click.

Setting up the Lancing Device

6

Penetration 

depth

Release Button

Cap

Number 

of lancets 

remaining

 WarNiNG

You must not insert the lancet 

drum into the lancing device and 

simultaneously press the release 

button or hold the lancing device 

with the release button resting 

on a surface such as a table top. 

This could release a lancet and 

inadvertently cause injury.

1 Lancet Drum = 6 Lancets

When ready to test…

Advance to next lancet

Prepare the lancing device for the 

next test. Slide the lever over and 

back once to advance to the next 

lancet. The window shows the 

number of lancets remaining.

Before you finish…

This file may not print or view at 100%.
Die lines and color breaks do not print.

Roche USA –51825 
V4/1 –07054238001_01 – Black

Roche USA –51825 
V4/2 –07054238001_01 – Cyan

Roche USA –51825 
V4/3 –07054238001_01 – Magenta

Roche USA –51825 
V4/4 –07054238001_01 – Yellow

51825_07054238001_01.indd   1

9/9/13   1:07 PM

Reviews: