The maximum lengths of the connecting cables are the following:
-
power supply device: 30 m (98 ft)
-
analog inputs: 100 m (328 ft)
-
power supply 0-20 mA/4-20 mA/0-5 V ratiometric/0-10 V
transducers: 100 m (328 ft)
-
free of voltage digital inputs: 100 m (328 ft)
-
high voltage digital input: 100 m (328 ft)
-
digital output: 100 m (328 ft)
-
bipolar stepper motor output: 5 m (16 ft; 10 m (32 ft) with
shielded cable)
-
CAN port:
-
1,000
m
(3,280
ft)
with
baud
rate
20,000 baud
-
500 m (1,640 ft) with baud rate 50,000 baud
-
250 m (820 ft) with baud rate 125,000 baud
-
50 m (164 ft) with baud rate 500,000 baud
-
power supply remote user interface: 30 m (98 ft)
-
RS-485 port: 1,000 m (3,280 ft); also look at the
MODBUS
specifications and implementation guides
manual.
One suggests using the connecting kit CJAV17 (female extractable
screw terminal blocks pitch 5.0 mm (0.196 in), to order
separately).
Operating temperature:
from -10 to 60 °C (from 14 to 140 °F).
Storage temperature:
from -20 to 70 °C (from -4 to 158 °F).
Operating humidity:
from 10 to 90% of relative humidity not condensing.
Pollution situation:
2.
Power supply:
24 VAC +10% -15%, 50/60 Hz ±3 Hz, 40 VA max. not isolated or
24... 37 VDC, 22 W max. not isolated, supplied by a class 2 circuit.
If the device is powered in direct current, it is necessary to respect
the polarity of the power supply voltage.
Protect the power supply with a fuse rated 2 A-T 250 V.
Overvoltage category:
III.