BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
5
–
9
4. Changing the sensor type will not change any other sensor constraint.
Table 5-1
Holding registers (Continued)
Register
Name
Address Size
Variable
Type
Description
Write
Access
4-20mA
Input
Calibration
Values
5667
8
Float
Two calibration values for a
low point and a high point,
each with reference value
and device output. Refer to
Yes
Measure
-
ment Values
4895
14
Mixed
14 for
details.
No
Sensor
Constant f1
4911
2
Float
Sensor constant f1.
Allowed range: 0.000 to
9.999
Yes
Sensor
Constant
dPhi1
4913
2
Float
Sensor constant dPhi1.
Allowed range: -9.99999 to
+9.99999
Yes
Sensor
Constant
dPhi2
4917
2
Float
Sensor constant dPhi2.
Allowed range: -9.99999 to
+9.99999
Yes
Sensor
Constant
dKSV1
4919
2
Float
Sensor constant dKSV1.
Allowed range: -9.99999 to
+9.99999
Yes
Sensor
constant
DKSV2
4921
2
Float
Sensor constant dKSV2.
Allowed range: -9.99999 to
+9.99999
Yes
Sensor
Constant m
4923
2
Float
Sensor constant m.
Allowed range: 0.00 to
+999.99
Yes
Sensor Type
4925
2
Int32
Sensor Type where:
0x00 = OP-3
0x01 = OP-6
0x02 = OP-9
4
Yes
Manual
Temperature
Compensa
-
tion
5611
1
Boolean
Activates the temperature
measurement of the Pt100
sensor by setting and uses
the manual temperature
value by deleting this bool
-
ean register. After writing
this register, the manual
temperature value has to
be set (register 2411).
Yes