OXY5500 Optical Oxygen Analyzer
BA02195C
4
–
60
Hauser
3.
Press the Set button next to the
1st Point
when the reading is steady.
The last measured value will be displayed in the top row next to the
selected port.
4.
Apply a higher value to the OXY5500.
5.
Enter this value in the
Reference
column in the
2nd Point
row.
6.
Press the Set button next to the
2nd Point
when the reading is
steady.
7.
The row
Test Point
displays the calibrated value, which is used for
calculating the pressure value. This value should match the
reference device value by less than 0.05mA.
This value is the uncalibrated value that will be used as the 1st
Point calibration value.
This value is the uncalibrated value that will be used as the 2nd
Point calibration value.
Figure 4–60
Analogues - 4-20mA input
calibration