17
Ethernet IP
Example:
Writing the following bytes:
0x 01 10 16 2D 00 08 10 00 C0 00 00 00 A8 00 00 00 01 00 00 00 0A 00 00 1F B1
will result in the IP 192.168.1.10
In detail:
0x 01
Slave address (int32 “01”)
0x 10
Function Code
0x 16 2D
Starting Address (5677 in int32 representation)
0x 00 08
Quantity of registers
0x 00 C0 00 00
Octet 1 (int32 192)
0x 00 A8 00 00
Octet 2 (int32 168)
0x 00 01 00 00
Octet 3 (int32 1)
0x 00 0A 00 00
Octet 4 (int32 10)
0x 1F B1
CRC16
Hauser recommends deactivating the current
measurement before changing any settings. The device will hold
its last analog output value until the next measurement.
Table 5–15
Definition of register 5677
Start
Register
Number of
Registers
Reg1/
Reg2
Reg3/
Reg4
Reg5/
Reg6
Reg7/
Reg8
Write
Access
5677
8
Int32:
Ethernet IP
Octet 1
Int32:
Ethernet IP
Octet 2
Int32:
Ethernet IP
Octet 3
Int32:
Ethernet IP
Octet 4
Yes