BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
4
–
45
Setting the first calibration point
Cal0
1.
Flow Cal0 gas to sensor for the first calibration point. Refer to
Table 4–5 for Cal 0 gas specifications.
2.
Press the Start button to the left of the
Cal0
value.
The
Status
field will display the message ‘Wait - Stabilizing!’ Wait for
the phase values to stabilize within
±
0.01
°.
3.
Run zero gas until the phase is stable; within 0.01 (approximately
45 to 60 minutes).
4.
Move to the Set button to the left of the
Cal0
value and press
.
Setting the second calibration point
Cal2nd
1.
Flow Cal2nd gas to sensor for the second calibration point.
2.
In the
O2-2nd
field, type in the oxygen value (concentration unit) of
the second calibration medium.
3.
Press the Start button next to the
Cal2nd
box.
The
Status
field will display the message, “Wait - Stabilizing!”. Wait
for the phase values to stabilize within
±
.01
°.
4.
Press the Start button to the left of the
Cal2nd
value.
5.
Press
.
Saving the calibration values
1.
Press the Save button at the upper right of the screen.
Disregard the “Ready to Set Value” message.
Refer to Table 4–5 on page 4–44 for Cal 2nd gas specifications.
Disregard the “Ready to Set Value” message.