OXY5500 Optical Oxygen Analyzer
BA02195C
4
–
20
Hauser
With Auto
selected, the measurement temperature is determined by the RTD
(Pt100) sensor.
Setting the temperature compensation
1.
Change the settings to the desired unit of measurement in the lower
right corner of the
Temperature
box. Figure 4–22 shows the
temperature set at 22.0
°
C.
OR
Select Manual if the temperature during measurement at the oxy
-
gen sensor is known and constant throughout the measurement.
a.
Switch to the desired temperature unit and change the
temperature value in the input field to the measurement
temperature.
Pressure compensation
On the MEASUREMENT SETTINGS screen, use the navigation buttons to move
to the
Pressure
box. Refer to Figure 4–23.
Automatically measured temperature values can be displayed in
°
C,
°
F or K.
Manual setting is only required if the temperature probe is not
19 before
using the Manual setting.
Temperature values can be input in
°
C,
°
F or K, in a range from
-99
°
C to 199
°
C. Values will automatically be recalculated in the
respective unit.
Summary of Contents for SpectraSensors OXY5500
Page 145: ......