OXY5500 Optical Oxygen Analyzer
BA02195C
3
–
8
Hauser
Analyzer Connections
The fiber optic oxygen cable to the SMA connector, located at the bottom of the
OXY5500, will be installed at the factory. Additional connectors are available as
shown in Figure 3–3.
Analog Outputs/Analog Input Connections
The OXY5500 is equipped with two independent analog outputs and one analog
input. The 4-20 mA current loop and serial output are connected terminal
blocks located inside the analyzer electronics enclosure. By default, the
4-20mA current loop analog outputs (IOUT1/IOUT2) are set to inactive.
RS-232/RS-485 Interface
: The unit has standard RS-232
communication via Modbus protocol. Use care in making
connections as described in
page 5-1 to avoid communication problems and potential
damage to the unit.
Optical Module with SMA Connector
: The optical module with
SMA connector is used to connect to the oxygen probe, which is
installed at the factory.
USB Connection
: The USB connection is for service and
troubleshooting purposes only. Do not connect during normal
operation. To avoid damage to the port, use only the USB Mini B
cable to connect to the unit. Refer to the Service Software
Operations Manual (P/N 4900002254) for system requirements.
Ethernet
: The unit uses standard Modbus - TCP/IP
communication. Use a CAT5 (or better) cable and make
connections per IEEE 802.3 standard.
OPTICAL MODULE WITH
SMA CONNECTOR
RJ-45
USB
Figure 3–3
Analyzer connections
FUSE BOX
TB2
TB1