OXY5500 Optical Oxygen Analyzer
BA02195C
5
–
20
Hauser
1-Point calibration of an OP-9 sensor
•
Precondition: Sensor is connected and put in a low oxygen
environment. Sensor constants are already set correctly (OP-9).
The goal for this example is to calibrate the oxygen sensor.
Table 5–19
Configuration for an analog output
Step
Description
Register(s)
Value
1
Deactivate the current measure
-
ment, otherwise the output may
generate false values.
5707, 5708
0 (Int32)
2
Set the Mode to “linear”.
4359, 4360
2 (Int32)
3
Set the Output to “oxygen”.
4363, 4364
1 (Int32)
4
Set the Error Level to “2mA”.
4389, 4390
2 (Int32)
5
Set the Low Level to “10.00”.
4377, 4378
10.00 (Float)
6
Set the High Level to “110.00”.
4381, 4382
110.00 (Float)
The oxygen value does not have to be set. This is done
automatically when setting the sensor type.
Table 5–20
1-Point calibration of a OP-9 sensor
Step
Description
Register(s)
Value
1
Read the current measurement
values.
4899 to 4908
Refer to
2
Check to confirm there are no
errors, especially error bits 1, 2, 4,
5 and 6. Refer to Table 5–10. Pro
-
ceed only when there are no errors
present.
3
Set the calibration values cal0 and
T0.
5521 to 5524
1st Float: 66.32
2nd Float: 21.98