![Endress+Hauser SpectraSensors OXY5500 Operating Instruction Download Page 111](http://html1.mh-extra.com/html/endress-hauser/spectrasensors-oxy5500/spectrasensors-oxy5500_operating-instruction_2407533111.webp)
BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
5
–
21
Table 5–21
Measurement reading for calibration process example
Register
4899/4900
Register
4901/4902
Register
4903/4904
Register
4905/4906
Register
4907/4908
Float:
Oxygen
Amplitude in
m
V
Float:
Oxygen
Phase shift in
degrees
Float:
Temperature
in
°
C
Float:
Calculated Oxy
-
gen Value in unit
Int32: Error
Register.
Refer to
50592.62
(Sensor and
environment
dependent
value)
66.32
(Sensor and
environment
dependent
value)
21.98
This value can
be ignored while
calibration
process takes
place.
0 (Error
code.
Should be 0
if a sensor is
connected)