BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
5
–
15
4-20mA port1 calibration values
All values are transmitted in mA.
Table 5–10
Error codes for the error register
Bit
2
N
Value
Error
0
1
No RTD (Pt100)
1
2
No sensor selected
2
4
Amplitude too low
3
8
SD card defect
4
16
Reference amplitude out of range
5
32
Photo diode saturated
6
64
Signal overflow
7
128
Signal overflow
8
256
Reserved
9
512
Critical error. Refer to
10
1024
No pressure sensor / pressure sensor out of range
11
2048
Reserved
12
4096
Storage space full
Table 5–11
Definition of register 4329
Start
Register
Number of
Registers
Reg1/
Reg2
Reg3/
Reg4
Reg5/
Reg6
Reg7/
Reg8
Write
Access
4329
8
Float:
Device
Value Low
Point
Float:
Reference
Value Low
Point
Float:
Device
Value
High
Point
Float:
Reference
Value
High Point
Yes