BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
B
–
13
9.
Turn the conduit bracket parallel to the panel and carefully
disengage the probe from the Tee union (panel side). Refer to
10.
Extend the probe conduit away from the panel and remove the
fittings from the probe tip (panel side). Refer to Figure B–15.
Take care not to disturb the temperature probe while removing
the oxygen probe conduit.
Take care to safely set aside the tube nut, cable gland cap and
plastic ferrules for use with the replacement probe.
Figure B–14
Remove probe from Tee union (panel side)
Figure B–15
Fittings on oxygen probe (panel side)
PLASTIC
FERRULE
TUBE NUT
CABLE GLAND
CAP
CABLE GLAND