recommends 35 days as a starting point for re-calibration or as the application
requires it. Refer to Table 4–2 below and
Logging and data management
On the MEASUREMENT SETTINGS screen, use the navigation buttons to move
to the
Logging
box. Refer to Figure 4–25.
•
Select Off if you do not want to store measurement data.
•
Select On to store measurement data.
The screen will switch to the
Measurement Browser
automatically. A
list will display showing the
Measurement
file name, the number of
measurement
Points
stored in the respective file and the date the file
was
Last Used
. Refer to Figure 4–26.
Table 4–2
Interval sampling rate/calibration frequency
Sampling Rate
Points
Calibration
Frequency (Days)
30 seconds
100,000
34.7
1 minute
100,000
69.4
1 hour
100,000
4,166
10 hours
100,000
41,666
In the Status bar, the
symbol indicates Logging is turned
off as shown in Figure 4–25.
Figure 4–25
Measurement settings screen -
Logging