BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
3
–
3
Tools:
•
Drill and bits
•
Tape measure
•
Level
•
Pencil
•
Screw driver (Philips)
•
Screw driver, small (Flat-head)
•
Needle-nose pliers
Analyzer Mounting
The OXY5500 analyzer is manufactured for wall or Unistrut
®
(or equivalent)
metal framing installations. Depending on your application and configuration,
the analyzer will come mounted on a plate or Unistrut frame. Refer to
Appendix A for drawings with detailed wall mounting dimensions.
When mounting the analyzer, be sure not to position the
instrument so that it is difficult to operate adjacent devices. Allow
1 m (3 feet) of room in front of the analyzer and any switches.
It is critical to mount the analyzer so that the supply and return
lines reach the supply and return connections on the chassis while
still maintaining flexibility so that the sample lines are not under
excessive stress.
Mounting brackets for equipment exceeding 18 kg intended for
wall mounting and/or part(s) that support heavy loads shall
withstand four times the maximum static load.
Because the breaker in the power distribution panel or switch will
be the primary means of disconnecting the power from the
analyzer, the power distribution panel should be located in close
proximity to the equipment and within easy reach of the operator,
or within 3 meters (10 feet) of the analyzer.