BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
3
–
11
5
GNDA
Analog output #1 ground
Configurable analog
output #1
6
IOUT1
Analog output #1 (4 – 20 mA); max
load = 800 Ohm
7
GNDA
Analog output #2 ground
Configurable analog
output #2
8
1
IOUT2
Analog output #2 (4 – 20 mA); max
load = 800 Ohm
9
NC
Not connected
—
10
Psense-
Analog input (4 – 20 mA); Sense (-)
Pressure sensor
input
11
Analog Input (4 – 20 mA); Sense (+)
loop-power 16 to 24 VDC; max
current = 32 mA
12
RTD +
4-wire RTD Pt100; Sense (+)
Temperature probe
13
RTD -
4-wire RTD Pt100; Sense (-)
Temperature probe
14
FRC+
4-wire RTD Pt100; Force (+)
15
FRC-
4-wire RTD Pt100; Force (-)
16
GNDT
RTD ground (Shield)
1. The 4-20 mA outputs are configured as sourcing to provide power to the loop. If a
PLC/HMI is used to provide power to the loop, an isolator will be required.
Table 3–2
Terminal block TB1
Pin
Label
Description
Function
1
V1+
Power supply 24 VDC - Factory
connection
DC power input
2
V2+
Power supply 24 VDC - Factory
connection
DC power input
3
GND
Power supply GND - Factory
connection
Power ground
4
GND
Power supply GND - Factory
connection
Power ground
5
232TX
RS-232 transmitter output
(typical signal level ± 6 V)
RS-232 signal
transmit
6
232Rx
RS-232 receiver input
(typical signal level ± 6 V)
RS-232 signal receive
Table 3–1
Terminal block TB2 (Continued)
Pin
Label
Description
Function