OXY5500 Optical Oxygen Analyzer
BA02195C
3
–
2
Hauser
Lifting/carrying the analyzer
At approximately 5.44 Kg (12 lbs) without the sample conditioning system, the
OXY5500 can easily be lifted from the packaging and moved to the installation
location. Take care to lift or carry the analyzer by the enclosure and not by any
ancillary probes or cables or damage may occur.
If the analyzer is configured with an optional integrated sample conditioning
system (SCS), two individuals may be required to lift and move the analyzer
system. Refer to the OXY5500 SCS Operating Instruction (P/N BA02196C) for
more information.
Installing the Analyzer
Installing the analyzer is relatively easy requiring only a few steps that, when
carefully followed, will ensure proper mounting and connection. This section
includes information regarding:
•
Hardware and Tools for Installation
•
•
Connecting Electrical Power to the Analyzer
•
Analog Outputs/Analog Input Connections
Required Basic Equipment
The following components are shipped with the OXY5500 analyzer from the
factory for installation and operation:
•
Flow-through Tee fitting with probe
•
Flow-through Tee fitting for temperature probe and pressure sensor
(pressure sensor is optional)
Hardware and Tools for Installation
Depending on the particular configuration of accessories and options ordered,
you may need the following hardware and tools to complete the installation
process.
Hardware:
•
1/4 in. (~6 mm) thickness Unistrut
®
(or equivalent) bolts and spring
nuts
•
Stainless steel tubing (using 1/4 in. [~6 mm] O.D. x 0.035 in. wall
thickness, seamless stainless steel tubing is recommended)
•
3/4 in. conduit or appropriate Ex e M20 cable gland
•
1/4 in. (M6) screws with an appropriate screw length for wall
material, e.g., concrete, drywall, etc.