OXY5500 Optical Oxygen Analyzer
BA02195C
4
–
44
Hauser
3.
Confirm settings for the specified probe being used.
4.
Allow nitrogen
(N
2
) Cal 0 gas to flow through the system for 45 to
60 minutes to purge the system. Refer to Table 4–5.
In the upper section of the main screen, the Present Values measured by the
OXY5500 are displayed. Refer to Figure 4–45.
Settings specified in the calibration certificate must be used for
probes. Refer to the calibration certificate example in Figure 4–34
on page 4–32 and the analyzer screen shown in Figure 4–31 on
Table 4–5
Calibration gas specifications
Sensor Type
OP-3
OP-6
OP-9
Cal 0
Calibration with oxy
-
gen-free environ
-
ment (e.g.,
nitrogen).
Calibration in oxy
-
gen-free environ
-
ment (nitrogen).
Calibration in oxy
-
gen-free environ
-
ment (99.9999%
nitrogen).
Cal 2nd
Calibration value
optimal at 20.9% O
2
in N
2
(or ambient
air).
Calibration value
optimal between 1%
and 2% oxygen.
Calibration value
optimal between
100 to 200 ppm O
2
in N
2
.
Storage
Stability
2 years provided the sensor material is stored in the original
packaging
Figure 4–45
Calibration screen