: Temperature at the first calibration point.
– Select Auto to measure the temperature at the first calibration
point with the RTD probe (Pt100 temperature sensor).
– Select Manual if the first calibration point is known and remains
constant through the calibration process. Temperature values can
be input in
°
C,
°
F or K. Switch to the desired temperature unit
and change the temperature value in the input field.
•
T2nd
: Temperature at the second calibration point.
– Select
Auto at the first calibration point for automatic
temperature measurement.
– Select Manual to insert the changes to the calibration
temperature manually.
To proceed with the calibration, press the Next button at the top right of the
screen followed by the
.
Before starting the measurement, the OXY5500 must be calibrated. Refer to
the section
Calibrating the analyzer
Complete the calibration procedures in this section before starting the
measurement. First, refer to the required equipment and materials list in
Table 4–4. Figure 4–38 shows an illustration of the components to be used for
the cylinder regulator purge process.
Figure 4–37
Calibration temperature screen