4-20mA port2 calibration values
All values are transmitted in mA.
4-20mA input calibration values
All values are transmitted in mA.
Analog input and output value ranges
The values that define the range of the analog outputs/inputs (Low, Mid and
High Values of Analog Port1 and 2, and Analog Input) always use the units
shown in Table 5–14.
Table 5–12
Definition of register 4979
Start
Register
Number of
Registers
Reg1/
Reg2
Reg3/
Reg4
Reg5/
Reg6
Reg7/
Reg8
Write
Access
4979
8
Float:
Device
Value
Low
Point
Float:
Reference
Value Low
Point
Float:
Device
Value
High
Point
Float:
Reference
Value High
Point
Yes
Table 5–13
Definition of register 5667
Start
Register
Number of
Registers
Reg1/
Reg2
Reg3/
Reg4
Reg5/
Reg6
Reg7/
Reg8
Write
Access
5667
8
Float:
Device
Value
Low
Point
Float:
Reference
Value Low
Point
Float:
Device
Value
High
Point
Float:
Reference
Value High
Point
Yes
Table 5–14
Oxygen units for various output, sensor and
measurement mode configuration
Output
Unit
Sensor/Condition
Oxygen
% O
2
OP-3
Oxygen
% O
2
/ppm gas
OP-6
Oxygen
ppm gas
OP-9
Temperature
°
C
Always
Pressure
Mbar
Always