Diagnostics and troubleshooting
Prosonic S FMU95 PROFIBUS DP
64
Hauser
•
A 0x 231
Distance: BAD
Temperature: GOOD
•
A 0x 281
Distance: BAD
Temperature: BAD
•
W 0x 281
Distance: UNCERTAIN
Temperature: UNCERTAIN
•
W 0x 501
Distance: BAD
Temperature: BAD
•
A 0x 502
Distance: BAD
Temperature: BAD
•
W 0x 521
Distance: UNCERTAIN
Temperature: BAD
•
A 0x 641
Distance: BAD
Temperature: GOOD
•
A 0x 651
Distance: BAD
Temperature: GOOD
•
W 0x 651
Distance: UNCERTAIN
Temperature: GOOD
•
A 0x 661
Distance: BAD
Temperature: GOOD
•
W 0x 661
Distance: UNCERTAIN
Temperature: GOOD
•
W 0x 691
Distance: UNCERTAIN
Temperature: GOOD
•
W 0x 802
Distance: UNCERTAIN
Temperature: GOOD
Error in the Level Block (LE)
•
W 0x 601
Level: BAD
•
A 0x 604
Level: BAD
•
W 0x 611
Level: BAD
•
A 0x 671
Level: BAD
•
W 0x 801
Level: UNCERTAIN
Error in the Calculation Blocks (SL, AL, DL, LD, SF, AF, DF, FD)
A 00 820-832
Total: BAD
Mean: BAD