Prosonic S FMU95 PROFIBUS DP
Commissioning
Hauser
53
9.3.9
Parameter set "LVL N dist. map."
Navigation
Level → Level (LVL) N → Basic setup → LVL N dist. map.
Parameter
•
Actual distance
Displays the distance D currently measured between the sensor membrane and the
surface of the product.
•
Range of mapping
Specifies the range, starting from the sensor membrane, in which mapping is performed.
• If
Check distance
=
Distance = Ok
or
Dist. too small
:
Confirm the preset value.
• If
Check distance
=
Manual
:
Enter the desired mapping range.
•
Start mapping
Select
Yes
to start recording the mapping curve.
→ The
LVL N status
parameter set appears.
→ If the displayed distance is still too small: keep recording mapping curves until the
displayed distance and the actual distance match.
•
Status
Specifies the mapping status:
•
Enable map
The mapping curve is taken into consideration during signal evaluation.
•
Disable map
The mapping curve is not taken into consideration during signal evaluation but
remains stored in the device.
•
Delete map
The mapping curve is deleted.
9.3.10 Parameter set "US sensor N"
For multi-channel devices: disable unused sensor inputs in the
US sensor N
parameter set.
Navigation
Sensor management → FDU sensor N → Sensor operation
Parameter "Sensor operation"
Switch sensor N on or off.
9.4
Advanced settings
9.4.1
Configuring the local display
Type of visualization
A0036764
44
"Type" = "1x value+bargr."