Commissioning
Prosonic S FMU95 PROFIBUS DP
58
Hauser
9.5
Simulation
9.5.1
Level or volume simulation
Navigation
Level → Level N → Simulation
Parameter
•
Simulation
Select the variable to be simulated (level or volume)
•
Sim. level value
Only displayed if
Simulation
=
Sim. level
Enter the level to be simulated. The linearization and the output signal follow this value.
•
Sim. volume value
Only displayed if
Simulation
=
Sim. volume
Enter the volume to be simulated. The output signal follows this value.
9.6
Protecting settings from unauthorized access
9.6.1
Software locking
Locking
1. Navigate to
Device properties → Password/reset → Code
.
2. Enter a number ≠ 2
457.
The device is locked; no entries can be made.
Unlocking
‣
If an attempt is made to change a parameter, the device jumps to
Password/reset
.
Enter "
2
457
".
Entries can be made again.
9.6.2
Keypad lock
Locking
‣
Press all three keys simultaneously.
The device is locked; no entries can be made. The
symbol appears on the
display.
Unlocking
‣
If an attempt is made to change a parameter, the device jumps to
Password/reset
.
Key-locked
is displayed in the
Status
parameter.
Press all three keys simultaneously.
Entries can be made again.