Electrical connection
Prosonic S FMU95 PROFIBUS DP
22
Hauser
Cable entries
Precut openings on the bottom of the housing for the following cable entries:
• M20x1.5 (10 openings)
• M16x1.5 (5 openings)
• M25x1.5 (1 opening)
Use a suitable tool to cut out the openings.
6.2.2
Terminal compartment of aluminum field housing
Valid for:
Order code 030 (housing, material)
Option 3 (aluminum field mounting, IP66 NEMA4x)
L
WARNING
To ensure explosion protection:
‣
Ensure that all the terminals are located in the field housing. (Exception: terminal for
protective ground on the outside of the field housing).
‣
Connect the housing to the local ground potential (PML).
‣
To route the cables, only use cable glands that meet the explosion protection
requirements at the place of operation.
Access to terminal compartment
Mad
D-796
Pro
Ta >
Xaxx
Dat./
X =
1
Orde
Ser
.-
5
2
3
4
1
A0033256
18
Access to terminal compartment in aluminum field housing
1
Aluminum field housing, open
2
Nameplate
3
Terminal for protective ground
4
Display and operating module
5
Aluminum field housing, closed
Cable entries
• 12 M20x1.5 openings for cable entries are located on the bottom of the field housing.
• To establish the electrical connection: guide the cables through the cable entries and into
the housing. The electrical connection is then established in the same way as in the DIN
rail housing.
6.2.3
Terminal compartment of DIN rail housing
Valid for:
Order code 030 (housing, material)
Option 2 (DIN rail mounting PBT, IP20)