Prosonic S FMU95 PROFIBUS DP
Diagnostics and troubleshooting
Hauser
65
10.2.5 Error codes
Meaning of the error codes
•
Digit 1:
Type of error
• A: alarm
• W: warning
• E: error (the error behavior is defined by the user.)
•
Digits 2 and 3:
Input or output channel
"
00
" means that the error does not refer to a specific channel.
•
Digits 4 to 6:
Error code according to the following table
List of error codes
•
A 00 100
Software version does not suit hardware version.
•
A 00 101
Checksum error
→ Total reset and recalibration
•
A 00 102
Checksum error
→ Total reset and recalibration
•
W 00 103
Initializing - please wait
→ If the message does not disappear after a few seconds, replace the electronics.
•
A 00 106
Downloading - please wait
→ Wait for the download to finish.
•
A 00 110
Checksum error
→ Total reset and recalibration
•
A 00 111/112/114/115
Electronics defect
• → Switch off device and switch it on again.
• → If the error persists: call Hauser Service.
•
A 00 116
Download error
→ Repeat download.
•
A 00 117
Hardware after change not identified
•
A 00 125
Electronics defect
→ Replace electronics.
•
A 00 152
Checksum error
→ Perform total reset and recalibration.
•
W 00 153
Initialization
→ If the message does not disappear after a few seconds, replace the electronics.
•
A 00 155
Electronics defect
→ Replace electronics.
•
A 00 164
Electronics defect
Replace electronics.