Description of device parameters
Proline Cubemass 300 Modbus RS485
88
Hauser
Selection
• Off
• Fixed value
• External value
• Current input 1
• Current input 2
Factory setting
Off
Additional information
Selection
• Fixed value
A fixed pressure value is used for compensation:
Pressure value
•
Current input 1
option,
Current input 2
option
The pressure value read in via the current input is used for compensation.
Pressure value
Navigation
Expert → Sensor → External comp. → Pressure value (6059)
Prerequisite
The
Fixed value
option is selected in the
Pressure compen.
Description
Use this function to enter a value for the process pressure that is used for pressure
correction.
User entry
Positive floating-point number
Factory setting
0 bar
Additional information
User entry
The unit is taken from the
Pressure unit
External press.
Navigation
Expert → Sensor → External comp. → External press. (6209)
Prerequisite
The
External value
option is selected in the
Pressure compen.
Description
Use this function to enter an external pressure value.
User entry
Positive floating-point number
Factory setting
0 bar
Additional information
User entry
The unit is taken from the
Pressure unit
*
Visibility depends on order options or device settings