Description of device parameters
Proline Cubemass 300 Modbus RS485
32
Hauser
Alarm delay
Navigation
Expert → System → Diagn. handling → Alarm delay (0651)
Description
Use this function to enter the time interval until the device generates a diagnostic
message.
The diagnostic message is reset without a time delay.
User entry
0 to 60 s
Factory setting
0 s
Additional information
Result
This setting affects the following diagnostic messages:
• 046 Sensor limit
• 140 Sensor sig.asym.
• 144 MeasErrorTooHigh
• 830 Sensor temp.
• 831 Sensor temp.
• 832 Electronic temp.
• 833 Electronic temp.
• 834 Process temp.
• 835 Process temp.
• 843 Process limit
• 862 Partly filled
• 912 Medium inhomog.
• 913 Medium unsuitab.
• 944 MonitoringFailed
"Diagn. behavior" submenu
Each item of diagnostic information is assigned a specific diagnostic behavior at the
factory. The user can change this assignment for specific diagnostic information in the
Diagn. behavior
submenu (→ 32).
The following options are available in the
Assign behavior of diagnostic no. xxx
parameters:
Options
Description
Alarm
The device stops measurement. The measured value output via Modbus RS485 and the
totalizers assume the defined alarm condition. A diagnostic message is generated.
The background lighting changes to red.
Warning
The device continues to measure. The measured value output via Modbus RS485 and the
totalizers are not affected. A diagnostic message is generated.
Logbook only
The device continues to measure. The diagnostic message is displayed only in the
Event
logbook
Event list
submenu (→ 191)) and is not displayed in
alternation with the operational display.
Off
The diagnostic event is ignored, and no diagnostic message is generated or entered.
For a list of all the diagnostic events, see the Operating Instructions for the device