Description of device parameters
Proline Cubemass 300 Modbus RS485
108
Hauser
Val.stat.inp.
Navigation
Expert → Input → Status input 1 to n → Val.stat.inp. (1353–1 to n)
Description
Displays the current input signal level.
User interface
• High
• Low
Active level
Navigation
Expert → Input → Status input 1 to n → Active level (1351–1 to n)
Description
Use this function to determine the input signal level at which the assigned function is
activated.
Selection
• High
• Low
Factory setting
High
Response time
Navigation
Expert → Input → Status input 1 to n → Response time (1354–1 to n)
Description
Use this function to enter the minimum time period for which the input signal level must
be present before the selected function is activated.
User entry
5 to 200 ms
Factory setting
50 ms
3.5
"Output" submenu
Navigation
Expert → Output
‣
Output
‣
Curr.output 1 to n
‣
PFS output 1 to n