Description of device parameters
Proline Cubemass 300 Modbus RS485
80
Hauser
Description
Use this function to enter a switch-on value for low flow cut off. Low flow cut off is
activated if the value entered is not equal to 0 → 80.
User entry
Positive floating-point number
Factory setting
Depends on country and nominal diameter→ 230
Additional information
Dependency
The unit depends on the process variable selected in the
Assign variable
parameter
Off value
Navigation
Expert → Sensor → Process param. → Low flow cut off → Off value (1804)
Prerequisite
One of the following options is selected in the
Assign variable
• Mass flow
• Volume flow
• Correct.vol.flow
Description
Use this function to enter a switch-off value for low flow cut off. The off value is entered as
a positive hysteresis from the on value→ 79.
User entry
0 to 100.0 %
Factory setting
50 %
Additional information
Example
1
A
Q
t
2
H
1
2
3
4
A
A0012887
Q
Flow
t
Time
H
Hysteresis
A
Low flow cut off active
1
Low flow cut off is activated
2
Low flow cut off is deactivated
3
On value entered
4
Off value entered