Proline Cubemass 300 Modbus RS485
Description of device parameters
Hauser
123
Additional information
Description
This setting does not affect the failsafe mode of other outputs and totalizers. This is
specified in separate parameters.
"Min." option
The current output adopts the value of the lower level for signal on alarm.
The signal on alarm level is defined via the
Current span
parameter (→ 111).
"Max." option
The current output adopts the value of the upper level for signal on alarm.
The signal on alarm level is defined via the
Current span
parameter (→ 111).
"Last valid value" option
The current output adopts the last measured value that was valid before the device alarm
occurred.
"Actual value" option
The current output adopts the measured value on the basis of the current flow
measurement; the device alarm is ignored.
"Defined value" option
The current output adopts a defined measured value.
The measured value is defined via the
Failure current
parameter (→ 123).
Failure current
Navigation
Expert → Output → Curr.output 1 to n → Failure current (0352–1 to n)
Prerequisite
The
Defined value
option is selected in the
Failure mode
Description
Use this function to enter a fixed value that the current output adopts in the event of a
device alarm.
User entry
0 to 22.5 mA
Factory setting
22.5 mA
Output curr. 1 to n
Navigation
Expert → Output → Curr.output 1 to n → Output curr. 1 to n (0361–1 to n)
Description
Displays the current value currently calculated for the current output.
User interface
3.59 to 22.5 mA