Error
no.
Display
Measures
Contact
Error current
Automatic
cleaning trigger
Fact.
User
Fact.
User
Fact.
User
E065
Current output range 2
exceeded
Check measured value and current
assignment.
yes
no
no
E066
Current output range 2
exceeded
yes
no
no
E067
Set value exceeded:
contr. / limit contactor 1
yes
no
no
E068
Set value exceeded:
contr. / limit contactor 2
yes
no
no
E069
Set value exceeded:
contr. / limit contactor 3
yes
no
no
E070
Set value exceeded:
contr. / limit contactor 4
yes
no
no
E071
Inaccurate measurement /
polarisation
Clean measuring cell; check table; choose
suitable measuring cell
yes
no
no
E077
Temperature outside
α
value table range
Clean measuring cell; check tables.
yes
no
no
E078
Temperature outside
concentration table
yes
no
no
E079
Conductivity outside
concentration table
yes
no
no
E080
Current output 1 param.
range too small
Spread current output.
no
no
—
—*
E081
Current output 2 param.
range too small
Spread current output.
no
no
—
—*
E100
Current simulation active
no
no
—
—*
E101
Service function yes
Switch service function off or
switch instrument off and back on.
no
no
—
—*
E102
Manual mode active
E106
Download yes
Wait for download to end.
no
no
—
—*
E116
Download error
Repeat download.
no
no
—
—*
E150
Distance between temp.
values in
α
value table too
small or not monotonically
increasing
Enter correct values in
α
value table
(minimum distance between temperature
values of 1 K required).
no
no
no
E151
USP error
no
no
no
E152
PCS alarm
Check measuring cell and connection.
no
no
no
E153
USP temperature error
no
no
no
* When this error is present, the cleaning function cannot be started.
(Field F8 does not exist for this error.)
LM253E07.CHP
Liquisys M CLM 223 / 253
Maintenance and troubleshooting
Hauser
57