Coding
Field
Selection or range
Factory setting
Display
Info
O211
Selection of
current range
4–20 mA
0–20 mA
O212
0/4 mA value;
enter
corresponding
measured value
cond/ind:
0.00 µS/cm
MOhm:
0.00 k
Ω⋅
cm
conc:
0.00 %
temp.:
0.0 °C
entire meas. range
Enter the measured value
corresponding to the minimum
current value (0/4 mA) at the
transmitter output.
Display format from A3.
(Spreading: see Technical data.)
O213
20 mA value;
enter
corresponding
measured value
cond/ind:
2000 mS/cm
MOhm:
500 k
Ω⋅
cm
conc:
9999 %
temp.:
150.0 °C
entire meas. range
Enter the measured value
corresponding to the maximum
current value (20 mA) at the
transmitter output.
Display format from A3.
(Spreading: see Technical data.)
O2 (2)
Current output
simulation
lin
= linear (1)
sim
=
simulation
(2)
table (3)
The simulation is terminated by
selecting (1) or (3).
See O2 (1), O2 (3) for other
characteristics.
O221
Enter simulation
value
current value
0.00 ... 22.00 mA
The current value entered here is
output through the current output.
O2 (3)
Enter current
output table
(S version only)
lin
= linear (1)
sim = simulation (2)
table
(3)
Values may also be added or
changed at a later point in time.
See O2 (1), O2 (2) for other
characteristics.
O231
Selection of table
option
read
edit
O232
Enter number of
table value pairs
1
1 ... 10
This is where the number of pairs
of x and y values (measured
value and associated current
value) is entered.
O233
Selection of table
value pair
1
1 ... number of table
value pairs
asign
Factory settings are printed in
bold
face;
base version does not include functions in italics.
O211
Sel.Range
4-20
O212
0/4 mA
0.00
µS/cm
O213
20 mA
2000
mS/cm
O2
Sel.Type
sim
O221
Simulat.
4.00
mA
O2
Sel.Type
table
O231
Sel.Table
read
O232
No.Elem.
1
O233
Sel.Elem.
1
LM253E5A.CHP
Liquisys M CLM 223 / 253
Instrument configuration
Hauser
27