7.4
Problem elimination based on error messages
Error
no.
Display
Measures
Contact
Error current
Automatic
cleaning trigger
Fact.
User
Fact.
User
Fact.
User
E001
EEPROM memory error
Switch instrument off and back on, return
instrument to your local Hauser
sales agency for repair or replace
instrument.
Load software compatible with hardware.
Load instrument software specific to
parameter measured.
yes
no
—
—*
E002
Instrument not calibrated,
calibration data invalid,
no user data or user data
invalid
(EEPROM error)
Software does not match
hardware (central module)
yes
no
—
—*
E007
Transmitter malfunction
Software does not match
hardware (transmitter)
yes
no
—
—*
E008
Measuring cell or
measuring cell
connection faulty
Check measuring cell and measuring cell
connection (E+H Service).
yes
no
no
E010
No temperature sensor
connected or temperature
sensor short-circuited
(temperature sensor
faulty)
Check temperature sensor and
connections;
if necessary, check instrument with
temperature simulator.
yes
no
no
E025
Limit for Airset offset
exceeded
Repeat Airset (in air) or replace measuring
cell.
Dry cell.
yes
no
no
E036
Calibration range of
measuring cell exceeded
Clean measuring cell and recalibrate; if
necessary, check measuring cell and
connections.
yes
no
no
E037
Below calibration range of
measuring cell
yes
no
no
E045
Calibration aborted
Recalibrate.
yes
no
—
—*
E046
Current output 1
parameter limits
interchanged
Set value for 20 mA > value for 4 mA.
yes
no
—
—*
E047
Current output 2
parameter limits
interchanged
Set value for 20 mA > value for 4 mA.
yes
no
—*
E049
Calibration range of
installation factor
exceeded
Check pipe diameter, clean measuring
cell and repeat calibration.
yes
no
—
—*
E050
Below calibration range of
installation factor
Check pipe diameter, clean measuring
cell and repeat calibration.
yes
no
—
—*
E055
Below measuring range of
main parameter
Immerse sensor in conductive medium or
perform Airset.
yes
no
no
E057
Measuring range of main
parameter exceeded
Check measurement, control and
connections.
yes
no
no
E059
Below temperature
measuring range
yes
no
no
E061
Temperature measuring
range exceeded
yes
no
no
E063
Below current output
range 1
Check configuration.
yes
no
no
E064
Current output range 1
exceeded
Check measured value and current
assignment.
yes
no
no