Problem
Possible cause
Remedy
Equipment needed, spare parts
Incorrect temperature
value
– Incorrect sensor connection
Verify connections using
connection diagram; three-wire
connection mandatory
Connection diagram in chpt. 3.4
– Measuring cable defective
Check cable for interruption/short/
shunt
Ohmmeter; also see chpt. 8.8.2 / .3
– Incorrect sensor type
Select sensor type on instrument
(field B1)
Incorrect conductivity
meas. value in process
– No / incorrect temperature
compensation
ATC: select compensation type;
linear: set correct
coefficient
MTC: set process temperature
– Incorrect temperature
measurement
Check temperature value
Reference instrument, thermometer
– Bubbles in medium
Suppress bubble formation:
- gas bubble trap
- counterpressure (cover)
- bypass measurement
– Polarisation effects (conductive
measuring cells only)
Use suitable measuring cell:
- use larger cell constant,
- use graphite in place of stainless
steel (resistance requirements)
Measuring range tables, e.g. in SI
“Conductivity”, or technical data of
conductivity measuring cells
– Flow rate too high (may cause
bubbles)
Reduce flow or choose
low-turbulence mounting position
– Potential in medium
(conductive only)
Ground medium close to meas. cell
Problem mainly occurs in plastic
lines and tanks
– Sensor soiled or coated
Clean sensor (s. chpt. 8.8.1)
Heavily soiled media: use spray
cleaning
Measured value
fluctuates
– Measuring cable interference
Connect cable screen acc. to
connection diagram
See chpt. 3.5
– Signal output line interference
Check line routing, try separate
line routing
Separate routing of signal output
and meas. input lines
– Interference potential in medium
Eliminate source of interference or
ground medium close to
conductivity measuring cell
Controller or timer
cannot be activated
– No relay module installed
Install LSR1-2 or LSR1-4 module
See chapters 8.4 and 8.5
Controller / limit contact
does not work
– Controller switched off
Activate controller
See chpt. 5.5 or fields R2xx
– Controller in “Manual / Off” mode
Choose “Auto” or “Manual / On”
mode
Keyboard, REL key
– Pickup delay setting too long
Disable or shorten pickup delay
See fields R2xx
– “Hold” function active
“Autom. hold” during calibration,
“hold” input activated;
“hold” via keyboard active
See fields S2 to S4
Controller / limit contact
works continuously
– Controller in “Manual / On” mode
Set controller to “Manual / Off” or
“Auto”
Keyboard, REL and AUTO keys
– Dropout delay setting too long
Shorten dropout delay
See fields R2xx
– Control loop interruption
Check measured value, current
output or relay contacts, actuators,
chemical supply
No conductivity current
output signal
– Line open or short-circuited
Disconnect line and measure
directly on instrument
mA meter 0–20 mA
– Output defective
See chpt. 8.3
Fixed current output
signal
– Current simulation active
Switch off simulation
See field O2
– Processor system out of sync
Switch instrument off and back on
EMC problem: check installation if
problem persists
Incorrect current output
signal
– Incorrect current assignment
Check current assignment:
0–20 mA or 4–20 mA?
Field O211
– Total load in current loop
excessive (> 500
Ω
)
Disconnect output and measure
directly on instrument
mA meter for 0–20 mA DC
Maintenance and troubleshooting
Liquisys M CLM 223 / 253