5
Instrument configuration
After power-up (connection to power), the
instrument performs a self-test and then
enters the measuring mode.
Now it can be configured and calibrated for
the first time.
The following function groups are available
on the Liquisys M (the groups that are only
available on the S version are marked
accordingly in the function descriptions):
Setup mode
➜
SETUP 1 (A)
see chpt. 5.2.1
➜
SETUP 2 (B)
see chpt. 5.2.2
➜
CURRENT OUTPUT (O) see chpt. 5.3
➜
ALARM (F)
see chpt. 5.4.1
➜
CHECK (P)
see chpt. 5.4.2
➜
RELAY (R)
see chpt. 5.5
➜
ALPHA TABLE (T)
see chpt. 5.6
➜
CONCENTRATION (K)
see chpt. 5.7
➜
SERVICE (S)
see chpt. 5.8
➜
E+H SERVICE (E)
see chpt. 5.9
➜
INTERFACE (I)
see chpt. 5.10
Calibration mode
➜
CALIBRATION (C)
see chpt. 5.11
Selecting and locating functions is facilitated
by a code displayed for each function in a
special display field. The structure of this
coding is shown in Fig. 5.2.
The first column indicates the function group
as a letter (see group designations). The
functions in the individual groups are counted
from the top to the bottom and from the left to
the right.
Factory settings
When the instrument is switched on for the
first time, the factory settings are in effect. The
following table provides an overview of all
major settings.
Please refer to the description of the
individual functions in chapter 5 for all other
factory settings (the factory settings are
printed in
bold
face).
2
s
F3
Err.Delay
DISPL4-E.CDR
Function coding:
The displayed code indicates
the function position in the
function group.
Additional information
Fig. 5.1
Function display:
auxiliary information for user
A
A1
A111
A121
A131
A132
A133
MATRIX.CDR
Fig. 5.2
Function coding
Type of measurement
Conductive or inductive conductivity measurement,
temperature measurement in °C oder °F (according to
instrument version ordered)
Temperature compensation type
Linear with reference temperature 25 °C
Temperature compensation
Automatic (ATC on)
Limit for controller 1
9999 mS/cm
Limit for controller 2
9999 mS/cm
Instrument configuration
Liquisys M CLM 223 / 253
22