5.2.2
Setup 2
Coding
Field
Selection or range
Factory setting
Display
Info
A4
Selection of unit to
be displayed
auto,
µS/cm, mS/cm,
S/cm, µS/m, mS/m,
S/m, auto
Ω
, k
Ω⋅
cm,
M
Ω⋅
cm, k
Ω⋅
m
When “auto” or “auto
Ω
” is
selected, the maximum
resolution possible is
automatically selected.
A4 not active if A1 = conc.
A5
Enter cell constant
for measuring cell
connected
cond:
1.000 cm
–1
ind:
1.98 cm
–1
MOhm:
0.01 cm
–1
0.0025 ... 99.99 cm
–1
Refer to the measuring cell
quality certificate for the exact
cell constant.
A6
Enter cable
resistance
0
Ω
0 ... 99.99
Ω
Only if A1 = cond
(conductive instrument).
See chapter 10.
A7
Enter measured
value damping
1
1 ... 60
Measured value damping causes
averaging over the specified
number of individual measured
values. It is used, for example, to
stabilise the display with
applications that fluctuate a great
deal. There is no damping if “1”
is entered.
Factory settings are printed in
bold
face;
base version does not include functions in italics.
Coding
Field
Selection or range
Factory setting
Display
Info
B
Function group
SETUP 2
Initial display in function group
SETUP 2.
B1
Selection of
temperature
sensor
Pt100
Pt1k = Pt 1000
NTC30
fixed
“fixed”:
No temperature measurement
if a fixed temperature value is
preset.
B2
Selection of
temperature
compensation
type
lin
=
linear
Tab = table
NaCl = common salt
(IEC 746)
Pure = ultrapure water
none
This option is not displayed for
concentration measurement.
“Pure” is only available on
conductive instruments.
auto
A4
Unit
1.000
A5
Cellconst
1/cm
A6
Ohm cable
0
@
1
A7
Damping
B
SETUP 2
B1
ProcTemp.
Pt100
B2
TempComp.
lin
LM253E5A.CHP
Liquisys M CLM 223 / 253
Instrument configuration
Hauser
25