Coding
Field
Selection or range
Factory setting
Display
Info
C126
Store calibration
results?
yes
no
new
If C125 = E xxx, then only no
or
new.
If new, return to C.
If yes/no, return to “Measurement”.
C1 (3)
Calibration
with measuring
cell adaptation for
inductive
measuring cells
(S version only)
Airs = Airset (1)
cellc = cell constant (2)
InstF
=
installation
factor
(3)
Measuring cell calibration with
compensation of wall influence.
The measuring cell remains at the place of installation.
C131
Enter calibration
temperature (MTC)
25.0 °C
–35.0 ... 250.0 °C
Only exists if B1 = fixed.
C132
α
value of
calibration solution
2.10 %/K
0.00 ... 20.00 %/K
This value is determined by the
calibration solution.
C133
Enter correct
conductivity value
of calibration
solution
current meas. value
0.0 ... 9999 mS/cm
The practical range is depends
on the measuring cell, i.e. the
calibration solution should be at
approx. 40 % of the measuring
range determined by the cell
(see chpt. 9, Fig. 9.2).
The display unit is always mS/cm.
C134
Calculated
installation factor
is displayed
1
0.10 ... 5.00
C135
Calibration status
is displayed
o.k.
E xxx
If the calibration status is not o.k.,
then the second display line
shows an explanation of the error.
C136
Store calibration
results?
yes
no
new
If C135 = E xxx, then only no
or
new.
If new, return to C.
If yes/no, return to “Measurement”.
Factory settings are printed in
bold
face;
base version does not include functions in italics.
yes
C126
Store
InstF
C1
Calibrat
25.0
C131
MTC temp.
°C
2.10
C132
alpha val
10.30
C133
Real val.
mS/cm
1
C134
InstFact
o.k.
C135
Status
yes
C136
Store
LM253E5D.CHP
Liquisys M CLM 223 / 253
Instrument configuration
Hauser
51