Coding
Field
Selection or range
Factory setting
Display
Info
T
Function group
ALPHA TABLE
(S version only)
T1
Selection of table
option
read
edit
T2
Enter number of
table value pairs
1
1 ... 10
Up to 10 value pairs can be
entered in the
α
table. These are
numbered from 1 ... 10 and can
be edited individually or in
sequence.
T3
Selection of table
value pair
1
1 ... number of table
value pairs
asign
T4
Enter temperature
value (x value)
0.0 °C
–35.0 ... 250.0 °C
The temperature values must
have a minimum distance of 1 K.
Defaults for the x values of the
table value pairs:
0.0 °C; 10.0 °C; 20.0 °C; 30.0 °C ...
T5
Enter temperature
coefficient
α
(y value)
2.10 %/K
0.00 ... 20.00 %/K
T6
Enter whether or
not the table
status is okay
yes
no
If “yes”, return to T.
If “no”, return to T3.
Factory settings are printed in
bold
face;
base version does not include functions in italics.
T
ALPHA TAB
T1
Sel.Table
read
T2
No.Elem.
1
T3
Sel.Elem.
1
T4
Temp.val.
0.0
°C
T5
Alpha val
2.10
%/K
T6
Status ok
yes