5.1
Start-up
After power-up (connection to power supply
established), the user must make the
following selections in the specified function
groups:
•
Function group SERVICE (S)
S1: Select language and exit function
group.
•
Function group SETUP 1 (A)
Adjust all the parameters in this group; see
chpt. 5.2.1.
•
Function group SETUP 2 (B)
Adjust all the parameters in this group; see
chpt. 5.2.2.
Other configuration options are explained in
the chapter to follow for each menu.
5.2
System configuration
The system is configured using the function
groups SETUP 1 and SETUP 2. The
measurement type and measuring cell are
selected here, and the settings for
temperature measurement are made.
All the parameters in these two function
groups are to be configured to avoid
measuring errors or failure to measure at all.
5.2.1
Setup 1
Coding
Field
Selection or range
Factory setting
Display
Info
A
Function group
SETUP 1
Initial display in function group
SETUP 1.
A1
Operating mode
selection
cond
=
conductive
ind
= inductive
MOhm = resistance
conc
=
concentration
Display varies according to
instrument version:
– cond/resistance/conc
– ind/conc
Any change in operating
mode causes an
automatic reset of
user settings.
A2
Selection of
concentration unit
to be displayed
%
ppm
mg/l
TDS
none
A2 only active if A1 = conc.
A3
Selection of
display format for
concentration unit
XX.xx
X.xxx
XXX.x
XXXX
A3 only active if A1 = conc.
Factory settings are printed in
bold
face;
base version does not include functions in italics.
A
SETUP 1
cond
A1
Oper.Mode
%
A2
Conc.Unit
XX.xx
A3
Format
Instrument configuration
Liquisys M CLM 223 / 253
24
Hauser