Coding
Field
Selection or range
Factory setting
Display
Info
R251
Switch function of
R2 (5) off or on
Off
On
R252
Selection of start
pulse
int
=
internal
(timer-contr.)
ext
= external
(digital input 2)
i+ext = intern. + extern.
i+stp = internal,
suppressed
by external
There is no real-time clock.
External suppression is required
for irregular time intervals (e.g.
weekends).
R253
Enter pre-rinse
time
20 s
0 ... 999 s
Water is used for rinsing.
R254
Enter cleaning
time
10 s
0 ... 999 s
Cleaning agent and water are
used for cleaning.
R255
Enter post-rinse
time
20 s
0 ... 999 s
Water is used for rinsing.
R256
Enter number of
repeat cycles
0
0 ... 5
R253 ... R255 is repeated.
R257
Enter pause time
360 min
1 ... 7200 min
The pause time is the time
between two cleaning cycles.
R258
Enter minimum
pause time
120 min
1 ... R357 min
The minimum pause time
prevents continuous cleaning
when the cleaning trigger is
present.
Factory settings are printed in
bold
face;
base version does not include functions in italics.
Off
R251
Function
int
R252
CleanTrig
R253
PreRinse
20
s
R254
CleanTime
10
s
R255
PostRinse
20
s
R256
Rep. Rate
0
360
R257
PauseTime
min
120
R258
Min.Pause
min
Instrument configuration
Liquisys M CLM 223 / 253
40
Hauser