Memory build-up
Navigation
Expert → System → External memory → Memory build-up
Direct access code: 140001-000
Description
"Stack memory": no more data can be stored once the data carrier is full.
"Ring memory": once the data carrier is full the oldest data are deleted so that new data can
be stored (First in first out (FIFO).
The "Ring memory" setting refers only to the automatic storage of measured values.
Manual save functions ("Operation -> SD card -> Update/Save measured values") are
not affected.
Options
Stack memory, Ring memory (FIFO)
"Ring memory" can be selected only if "Save as" is set to "protected format" (and not
"CSV").
Factory setting
Stack memory
Warning at
Navigation
Expert → System → External memory → Warning at
Direct access code: 140005-000
Description
Issues a warning before the data carrier is x% full.
A warning is indicated on the device and this is also stored in the event buffer.
A relay can also be switched.
Only for external SD card (does not apply to USB stick)!
User entry
0 to 99%
Factory setting
90
Switches relay
Navigation
Expert → System → External memory → Switches relay
Direct access code: 140006-000
Description
When the warning "Data carrier full" is displayed a relay can also be switched on.
Options
Not used, Relay x
All the available relays are displayed
Factory setting
Not used