2.6.1 Installation - X-STREAM
XEGK, X-STREAM
XEGP
Fig. 2-7:
Socket X1 - Analog Outputs & Digital Outputs 1-4
Signal inputs and outputs (basic configuration)
The number of signal outputs actually available varies according to the unit’s configuration.
Analog signals
Relay outputs
Analog signals and relay outputs are located on a shared 25-pin submin socket X1,
or on an optional terminals adaptor XSTA (
page 2-21; not for XEGK).
Specification of analog signal outputs:
4(0)–20 mA; burden: R
B
≤ 500 Ω
Specification of relay outputs 1–4:
Dry relay change-over contacts can be used as
NO or NC.
Electrical specification:
max. 30 VDC, 1 A, 30 W
Notes!
• Consider the installation notes in section 2.7.
• Technical data and installation information for further available interfaces are given in the
X-STREAM
Enhanced
instruction manual. Connector layouts are part of the appendix of this
manual.
Note!
Configuration of relay contacts
as per standard factory setting (NAMUR
status signals).
NAMUR status signals
are automatically configured Fail Safe!
Signal
Pin
Analog
outputs
Channel 1, (+) 4 (0)–20 mA
1
Channel 2, (+) 4 (0)–20 mA
2
Channel 3, (+) 4 (0)–20 mA
3
Channel 4, (+) 4 (0)–20 mA
4
Channel 5, (+) 4 (0)–20 mA
5
unused
6
unused
7
Relay contacts (Status signals)
Output1 (Failure) NC
8
Output1 (Failure) NO
9
Output1 (Failure) COM
10
Output3 (Off spec) NC
11
Output3 (Off spec) NO
12
Output3 (Off spec) COM
13
Pin
Signal
14
Channel 1, GND
Analog
outputs
15
Channel 2, GND
16
Channel 3, GND
17
Channel 4, GND
18
Channel 5, GND
19
unused
20
Output2 (Maintenance request) NC
Relay contacts (Status signals)
21
Output2 (Maintenance request) NO
22
Output2 (Maintenance request) COM
23
Output4 (Function check) NC
24
Output4 (Function check) NO
25
Output4 (Function check) COM