Signal damping (set in DAMPING) allows
smoothing the output signal, but also affects
the response time of outputs and display:
• The factory setting is 0 seconds.
•
The maximum possible t
90
time is limited by
the size of the internal sampling buffer and
the sampling rates of the installed measur-
ing principles/sensors.
• Multi-channel units: the value for each
channel must be entered separately.
The last line of the second page (Pressure
Compensation) enables the user
•
to enter the current ambient pressure
manually, if no pressure sensor is in-
stalled,
or
• to view the current pressure, if a sensor
is installed (
INSTALLED OPTIONS).
If no pressure sensor is installed, enter the
current ambient pressure here and adjust
it, when significant changes take place: this
improves the accuracy of the instrument.
Ch1
Measurement 1of2
Ranges..
Damping..
Linearization..
Cross Interference..
Average..
Delay..
Ch1
Measurement 2of2
Cut-off:
None
Pressure Compensation ..
Ch1
Pressure compensation
Manual Pressure:
1013 hPa
Pressure
1013 hPa
Pressure Status
Good
3.5.7 Checking the Settings
If "AfterCalCheck" is enabled (
On
), the cali-
bration procedure adds an additional phase
where it measures and stores the concentra-
tion value shortly after the calibration.
Press
left
several times to return to SETUP
and open MEASUREMENT.