X-STREAM Non-Incendive
Instruction Manual
HASXENE-IM-EX
05/2017
The first line gives you the choice to cancel
the procedure now.
Select the second line to
start the calibra-
tion
.
The next lines show
•
the calibration gas setup (here: required
zero gas concentration is 0.000 ppm),
• the currently measured gas concentration
•
and the current gas flow, if a flow sensor
is installed.
Ch1
Zero calibration
Cancel!
Start!
Zero gas
0.000 ppm
Concentration
0.000 ppm
Flow
0.00 l/min
Status..
Results..
Restore!
4.1 Perform a calibration
When finished press
left
several times to
return to
either
SELECT COMPONENT
(multi channel ana-
lyzer only), to perform a zero calibration for
another channel,
or
to CONTROL, where you may start a span
calibration. The procedure and screens look
similiar to those of a zero calibration: