General Operating Notes
GENERAL OPERATING NOTES
• The unit must be installed in a clean and dry area, protected from strong vibrations and frost
.
• The unit must not be exposed to direct sunlight and sources of heat. Admissible ambient
temperatures (see technical details) must be adhered to.
• Gas inlets and
outlets must not be interchanged. All gases must be conditioned, before sup
-
plied to the unit. When using this unit with corrosive sample gases, ensure that these gases
do not contain components harmful to the gas lines.
• Admissible gas pressure for sample and test gases is 1500 hPa.
• Exhaust lines must be laid inclined downwards, depressurized, protected from frost and
according to applicable regulations.
• If it is necessary to disconnect the gas lines, the unit’s gas connectors must be sealed with
PVC caps to avoid polluting the internal gas lines with condensate, dust, etc.
• To ensure electromagnetic compatibility (EMC), only shielded cables (supplied by us on re
-
quest, or of equivalent standard) may be used. The customer must ensure that the shielding
is correctly fitted. Shielding and terminal housing must be electrically connected; submin-D
plugs and sockets must be screwed to the unit.
• When using optional external adapters (submin-D to screw-type terminals), protection against
electromagnetic interference is not ensured (CE compliance pursuant to EMC guidelines).
In this case the customer or operating company functions as a system builder and must
therefore ensure and declare compliance with EMC guidelines.
HAZARD TO LIFE AND EXPLOSION HAZARD BY EXHAUST GASES
Exhaust gases may contain hydrocarbons and other toxic gases such as
carbon monoxide.
Faulty gas connections may lead to explosion and death.
Ensure that all gas connections are connected as labelled and airtight.