![Emerson Rosemount Analytical X-STREAM Enhanced Series Short Form Manual Download Page 40](http://html.mh-extra.com/html/emerson/rosemount-analytical-x-stream-enhanced-series/rosemount-analytical-x-stream-enhanced-series_short-form-manual_103709040.webp)
Emerson Process Management GmbH & Co. OHG
2-4
X-STREAM XE
Short Form Manual
HASXEE-SFM-HS
05/2017
2.3.2
Specific
Technical Data
2.3 Installation - Technical Data
XEGK
XEGP
XEXF
Temperatures
operational
0…+50 °C
32…122 °F
0 (-20)…+50 °C
32 (-4)…122 °F
storage
-20…+70 °C
-4…-158 °F
-20…+70 °C
-4…158 °F
Weight, max
8…12 kg
17.6…26.5 lb
12…16 kg
26.5…35.3 lb
XEF: 25 kg / 55.1 lb
XDF: 45 kg / 99.2 lb
IP or Type rating
IP 20
indoor use,
protected against dripping water and
direct sun light
IP 66 / Type 4X
outdoor use ,
protected against
direct sun light
Measurement channels
max. 3
max. 5
Gas connections
max number
8
max for purging
(separate / incl.)
2 incl.
1 separate,
1 incl.
1 separate,
1 incl.
material
PVDF; stainless steel (opt.)
stainless steel
sizes
6/4 mm;
1
⁄
4
"
Power supply unit
external
(option)
wide range,
internal
Power supply
nominal voltage
24 V
100–240 V
50 / 60 Hz
voltage range
10…
30 V
85–264 V
47–63 Hz
nominal input current
standard, max
2.5 A
1.3–0.7 A
1.3–0.7 A
XEF: 1.3–0.7 A
XDF: 1.5–0.8 A
w/ temperature control,
max
n.a.
3–1.5 A
XEF: 3–1.5 A
XDF: 5.5–3 A
Power input fuses
AC 230 V
T 3.15 A
5x20 mm
AC 230 V
T 4 A
5x20 mm
AC 230 V
T 4 A
5x20 mm
AC 230 V
T 6.3 A
5x20 mm