3.8 Checking the Settings
The most important settings have now been
checked and the unit is configured to suite
your needs.
A
backup copy of these configuration data
can now be saved.
Press
left
until SETUP and then open SAVE-
LOAD.
3.8.6 Backup the Settings
This menu gives you the choice, to either
• make a local
backup to a protected memory
area
• restore the factory default settings, or
• make a
backup to an external
USB device.
You may turn the alarm function
On
or
Off
separately for each channel ("
Alarm Monitor
").
It´s also possible to use some of the thresh
-
olds only. In this case, set the not used to a
level outside the range limits (for this, "
Lo"
and "
LoLo"
support entering negative values).
In case of an alarm, you may like to enter
the STATES
submenu, to check which one
is triggered.
Ch1
Concentration
Alarm Monitor:
On
LoLo Level:
50.000 ppm
Lo Level:
100.000 ppm
Hi Level:
400.000 ppm
HiHi Level:
600.000 ppm
Hysteresis:
10.000 ppm
States..
Save-Load
Local Backup..
Factory Defaults..
USB Backup..
USB Firmware Update..